Design: Aluminum housing featuring integrated PC module, large graphical LC-display with touch screen, 4-channel CO2/H2O gas analyzer, flow control, CO2 control (supplied via small cartridges or cylinders) and H2O control (for drying and humidifying). Pneumatic connectors for air inlet, measuring head and four vents. Sockets for cable connections with Measuring Head 3010-S, PC (USB interface), two auxiliaries, two Li-ion Batteries 3025-A (or AC Power Supply 3020-N or external DC) and one additional component
CO2/H2O Gas Analyzer
Design: 4-channel CO2/H2O absolute NDIR gas analyzer, separate cuvettes for CO2 and H2O
CO2 measurement: Simultaneous absolute and differential measurements, absolute range: 0 to 3000 ppm, cuvette length 20 cm, cuvette volume of one cell 6 cm3, gas-filled detector
H2O measurement: Simultaneous absolute and differential measurements, absolute range: 0 to 75000 ppm, cuvette length 20 cm, cuvette volume of one cell 6 cm3, pyroelectric detector (solid state)
Max. noise in absolute mode: <0.2 ppm CO2 and <30 ppm H2O
Resolution: 0.01 ppm CO2, 1 ppm H2O
Linearization:
CO2 absolute: Max. error between 0 and 600 ppm: 12 ppm, above 600 ppm: 2% of measured value
Deviation of differential CO2 zero with changing CO2: Typ. less than 0.5 ppm
H2O absolute: Max. error between 0 and 15000 ppm: 330 ppm, above 15000 ppm: 2% of measured value
Deviation of differential H2O zero with changing H2O: Typ. less than 150 ppm
Correction: Signal is corrected for temperature, pressure and H2O-effect on CO2 signal
Other Specifications of the Control Unit 3000-C
Pneumatic time delay: Delay between Analyzer and Standard Measuring Head 3010-S is 2.5 s at standard flow rate
Barometric air pressure measurement: Range 60 to 110 kPa, accuracy ±0.1%
Mass flow measurement: Thermal mass flow meter, range 0 to 1500 μmol s-1, accuracy ±1%
Integrated user interface: Panel PC 486 SX/33 MHz with transflective graphical LC-display 320 x 240 dots (effective display area 11.5 cm x 8.6 cm) with backlight and touch screen
Data storage capacity: 64 MB flash card
CO2 control: Integrated CO2-control via thermal valve, range 0 to 2000 ppm, CO2 supply via CO2 cartridges (8 g CO2, provide more than 48 h continuous supply at 350 ppm, reserve is indicated) or via CO2 cylinder with pressure reducer
H2O control: Integrated H2O control via step motor for humidifying and drying, range 0 to nearly 100% r.h. (noncondensing)
Measured and calculated parameters: CO2 absolute, CO2 difference, H2O absolute, H2O difference, flow, ambient pressure, 2x auxiliaries, cuvette temperature (upper and lower half), leaf temperature, ambient temperature, PAR in upper part of the cuvette, PAR in lower part of the cuvette, external PAR, impeller frequency, evaporation, VPD, H2O conductance, net photosynthesis, intercellular CO2 concentration. Recalculation of stored data is possible
PC interface: USB 2.0
Auxiliaries: Two analog inputs, range 0 to 4095 mV
Power supply: Field replaceable rechargeable Li-ion Battery 14.4 V/15 Ah 3025-A (two batteries
supplied), external 16 V DC, AC Power Supply 3020-N for laboratory operation
Operating time: 2 to 4 hours typ. with one Li-ion battery, 4 to 8 hours typ. with two Li-ion batteries
Operating temperature: -5 to 45 °C
Dimensions: 43 cm x 28 cm x 26 cm (L x W x H)
Weight: 11.1 kg (incl. Li-ion battery)
Design: DC power supply unit for laboratory use
Output voltage: 16 V DC
Output power: 135 W
Mains power supply: 100 to 240 V AC, 50/60 Hz
Operating temperature: 0 to 60 °C
Dimensions: 20 cm x 8 cm x 5 cm (L x W x H)
Weight: 1 kg
Design: High-performance, maintenance-free, rechargeable Li-ion battery with protection circuit, Perspex-aluminum housing
Nominal voltage: 14.4 V
Typical capacity: 15 Ah
Operating temperature: -10 to +60 °C (discharge) and 0 to 50 °C (charge)
Dimensions: 28 cm x 15.5 cm x 2.8 cm (L x W x H)
Weight: 1.7 kg
Design: Mains operated automatic Li-ion battery charger with two independent charge outputs, 2-line LC-display with backlight for each charge output
Charging voltage: 16.4 V max.
Charging current: 0 to 3 A
Recharging time: 6 hours for one or two Li-ion Batteries 3025-A
Mains power supply: 100 to 240 V AC, 50/60 Hz
Operating temperature: -5 to 45 °C
Dimensions: 20.5 cm x 19 cm x 9 cm (L x W x H)
Weight: 1.7 kg