JUNIOR-PAM

Version:

Product

Teaching Chlorophyll Fluorometer

JUNIOR-PAM (Blue Version - JUNIOR-PAM)

Modulated fluorescence excitation
Blue LED (wavelength of maximum emission: 445 nm). Modulation frequencies 5 to 25 Hz adjustable in increments of 5 Hz, and 100 Hz. Fluorescence at wavelengths greater than 630 nm is measured
Actinic light
Same LED as used for modulated light. Actinic PAR, 25 to 1500 μmol μmol m-2 s-1; maximum PAR of saturation pulses, 7000 μmol m-2 s-1. All data valid for 1 mm distance from the tip of the 50 cm JUNIOR-PAM light guide
Housing
Aluminum housing with USB type B socket, M8 4-pole socket to connect the JUNIOR-PAM monitoring leaf clip (accessory), port for JUNIOR-PAM light guide, and swivel-mounted sample support
Far-red light
LED with 745 nm maximum emission wavelength for selective excitation of photosystem I
Fluorescence detection
PIN photodiode protected by long-pass filter. Selective window amplifier to measure pulse-amplitude modulated (PAM) fluorescence
Communication
Standard USB communication via 1.2 m USB-cable type A-B
Power supply
5 V DC supplied by the USB-cable type A-B used for communication
Power consumption
100 mW at normal operation and 500 mW during saturation pulse
Dimensions
11.5 cm x 6.5 cm x 3 cm (L x W x H)
Weight
200 g
Operating temperature
0 to + 40 °C
Operating humidity range
35 to 85% RH (to avoid condensation)
Program

WinControl-3 System Control and Data Acquisition Program (Microsoft Windows 10 and 11) for operation of measuring system via PC, data acquisition and analysis. Not compatible with Windows 10 on ARM

Saturation Pulse Analysis

Measured: Ft, F0, FM, F, F0’ (also calculated), FM’. Depending on the leaf clip connected, the software can record PAR, temperature and also humidity. [In the case of the MINI-PAM-II clip humidity can be measured, which the clip of the JUNIOR-PAM cannot.]
Calculated: F0’ (also measured), FV/FM and Y(II) (maximum and effective photochemical yield of PS II, respectively), qL, qP, qN, NPQ, Y(NPQ), Y(NO) and ETR (electron transport rate)

Fitting Routines

Two routines for determination of the cardinal points α, Ik and ETRmax of light curves

Programmed Features

Automatic determination of signal offset for all light intensities and gain levels. Automatic calibration of internal PAR sensor against an external PAR sensor connected to the instrument

Computer Requirements

Processor: 0.8 GHz, RAM: 512 MB, screen resolution: 1024 x 600 pixels, interface: USB 2.0/3.0

Communication Protocol

USB

Light guide
50 cm x 1.5 mm (length x diameter) plastic fiber
Sample clips
Open leaf clip (angle between incident radiation from JUNIOR-PAM and leaf surface, 60°) and magnetic leaf clip (angle between incident radiation from JUNIOR-PAM and leaf surface, 90°)
Fluorescence standard
3 cm x 2 cm fluorescence foil
Polishing set
Set for fiber tips including various polishing pads
USB cable
A-B USB cable
Software on USB memory key
Including latest WinControl-3 software
Transport Case JUNIOR-T
Grey plastic box with handle. Interior: convoluted foam padding. Dimensions: 29 cm x 25 cm x 6 cm (L x W x H). Weight: 510 g
Design

Padded plastic case with handle

Dimensions

25 cm x 29 cm x 5 cm (L x W x H)

Weight

680 g

Accessories

Micro quantum sensor
Selective PAR measurement, 0 to 2500 μmol m-2 s-1 PAR
Thermocouple
: Ni-CrNi, -20 to +60 °C
Leaf temperature
-20 to +60 °C
Power supply
JUNIOR-PAM special socket
Cable length
100 cm
Dimensions
13 cm x 2.5 cm (max.) x 5.5 cm (max.) (L x W x H)
Weight
162 g (including cable)